Microprocessor UV-VIS Spectrophotometer(Double Beam) – Quantum Scientific http://quantumscientific.in A Leading Manufacturer , Exporter and Supplier of Scientific, Laboratory, Pharma & Medical Equipments. Mon, 11 Nov 2019 03:54:35 +0000 en-GB hourly 1 https://wordpress.org/?v=5.0.3 http://quantumscientific.in/wp-content/uploads/2015/10/favicon-50x50.png Microprocessor UV-VIS Spectrophotometer(Double Beam) – Quantum Scientific http://quantumscientific.in 32 32 Microprocessor UV-Vis Spectrophotometer(Double Beam) http://quantumscientific.in/product/microprocessor-uv-vis-spectrophotometerdouble-beam-2/ http://quantumscientific.in/product/microprocessor-uv-vis-spectrophotometerdouble-beam-2/#respond Wed, 08 Mar 2017 09:03:10 +0000 http://quantumscientific.in/?post_type=product&p=5689 Technical Specifications Display Graphic LCD (320*240 Dots) Lamps Deuterium Lamp & Tungsten Halogen Lamp Detector Silicon Photodiode Dimensions(L*W*H) 630 x 430 x 206 mm Net Weight 26 Kg To measure the Absorbance and Transmittance Quantitative 1. Coefficient Method 2. Standard Curve Up to 10 standard samples may be used to establish a curve. Four methods […]

The post Microprocessor UV-Vis Spectrophotometer(Double Beam) appeared first on Quantum Scientific.

]]>
Technical Specifications
Display Graphic LCD (320*240 Dots)
Lamps Deuterium Lamp & Tungsten Halogen Lamp
Detector Silicon Photodiode
Dimensions(L*W*H) 630 x 430 x 206 mm
Net Weight 26 Kg
To measure the Absorbance and Transmittance
Quantitative
1. Coefficient Method
2. Standard Curve
Up to 10 standard samples may be used to establish a curve.
Four methods for fitting a curve through the calibration points:
Linear fit, Linear fit through zero, Square fit and cubic fit.
DNA/Protein Test
Concentration and DNA purity and easily calculated:-Absorbance ratios 260 nm / 280 nm with optional subtracted absorbance at 320 nm. DNA concentration=62.9xA260-36.0 x A2B0 Protein concentration=1552xA260-757xA280.
Wavelength Scan
1. The wavelength scan intervals are 0.1, 0.2, 0.5, 1, 2, 5 nm
2. High, Medium and Low scan speeds are available. They vary from 100 to 3600 nm/min.
3. Wavelengths are scanned from high to low so that the instrument waits at high WL and it minimize the degradation of UV sensitive samples.
Kinetics
This mode may be used for time course scanning or reaction rate calculations. Abs vs time graphs is displayed on the screen in real time.

Wait time and measurement time up to 12 hours may be entered with time intervals of 0.5,1,2,5,10,30 seconds and 1 min. Post-run manipulation includes re-scaling, curve tracking and selection of the part of the curve required for the rate calculation. Rate is calculated using a linear regression algorithm before multiplying by the entered factor.

The post Microprocessor UV-Vis Spectrophotometer(Double Beam) appeared first on Quantum Scientific.

]]>
http://quantumscientific.in/product/microprocessor-uv-vis-spectrophotometerdouble-beam-2/feed/ 0
Microprocessor UV-VIS Spectrophotometer(Double Beam) http://quantumscientific.in/product/microprocessor-uv-vis-spectrophotometerdouble-beam/ http://quantumscientific.in/product/microprocessor-uv-vis-spectrophotometerdouble-beam/#respond Wed, 08 Mar 2017 09:01:43 +0000 http://quantumscientific.in/?post_type=product&p=5685 Salient Features Wide wavelength range, satisfying requirements of various fields Fully automated design, realizing the simplest measurement & satisfying the requirements of pharmacopoeia. Maximum of 9 wavelenths & 7 samples scan be measured at one time. Automatic change-over between W lamp & D 2 lamp Optimized optics and large scale integrated circuits design, light source […]

The post Microprocessor UV-VIS Spectrophotometer(Double Beam) appeared first on Quantum Scientific.

]]>
Salient Features
  • Wide wavelength range, satisfying requirements of various fields
  • Fully automated design, realizing the simplest measurement & satisfying the requirements of pharmacopoeia.
  • Maximum of 9 wavelenths & 7 samples scan be measured at one time.
  • Automatic change-over between W lamp & D 2 lamp
  • Optimized optics and large scale integrated circuits design, light source and receiver from world famous measurement all add up to high performance and reliability
  • Rich measurement methods:wavelength scan, time scan, multi-wavelenth determination, multi-order derivative determination, double-wavelength method and triple-wavelength method etc., meet different measurement requirement.
  • Automatic 10mm 8-cell holder
  • Data output can be obtained via a printer port and a RS-232 interface
  • Parameters and data can be saved for user’s convenience
  • PC controlled measurement can be achieved for more accurate and flexible requirement
Technical Specifications
Wavelength Range 190-1100 nm
Spectral Bandwidth 1 nm
Working Mode T,A,C,E
Optical System Double beam
Scanning Speed Fast / Medium / Slow
Photometric Range -0.3 – 3.5 A, 0 to 220 %T
Wavelength Accuracy +0.3 nm
Wavelength Reproducibility < 0.10 nm
Photometric Accuracy + 0.3% T (0-100% T)
+0.002 A (0-0.1 A)
Photometric Reproducibility < 0.15 % T
Stray Light < 0.1% T (Nal. 220nm, NaNO2 340 nm)
Baseline Flatness + 0.002 A
Stability < 0.001A/h (at 500 nm, after warming up)
Noise +0.001A (at 500 nm, after warming up)
Detector Dual Silicon Photodiode
Display 6 inches high light blue LCD
Power AC:220V + 10%, 50 Hz.
Dimensions 630 x 470 x 210 mm
Weight 26 Kg
Accessories
Glass Cells 16 No.
Quartz Cells 4 Nos.
Fuse 3 Nos.
Instrument Cover 1 No.
Software CD 1 No.
RS 232 Cable 1 No.