Specifications:
PARAMETER | SENSOR | RANGE | RESOLUTION | ACCURACY |
2 | Electrochemical | 0-25% | 0.001 | 0.005 |
CO | Electrochemical | 0-4000 PPM | 1 PPM | ± 2% |
CO (A) Sensor Comp. For H2 | Electrochemical | 0-4000 PPM | 1PPM | ± 2% |
CO(B) | Electrochemical | 0-10% | 0.0001 | ± 2% |
CO2 | Calculated | 0-25% | 0.001 | |
NO | Electrochemical | 0-1999 PPM | 1PPM | ± 2% |
NOX | Electrochemical | 0-1999 PPM | 1PPM | ± 2% |
NO2 | Electrochemical | 0-100 PPM | 1PPM | ± 2% |
SO2 | Electrochemical | 0-1999 PPM | 1PPM | ± 2% |
CxHx (Calibrated for methane) |
Catalytic | 0-100% LEL | 0.01 | ± 2% |
or 0-5% V | 0.001 | ± 2% | ||
Excess Air (A) | Calculated | 0- | 0.001 | |
Temp. Air (Ambient) | PT 100 | 0-99°C | 1°C | ± .5% |
Temp. Gas | Tc K | 0-1000°C | 1°C | ± 0.25% |
Temp. Differential | Tc K | 0-1000°C | 1°C | ± 0.25% |
Combustion Efficiency | Calculated | 0-99.9% | 0.001 | ± 0.1% |
- Complete Automatic operation: Complete Automatic operation including Calibration, “ON – OFF” & “SAMPLE” Keys only
- Self diagnostic check: Self diagnostic check & specific fault indication, Discrete LEDs for Sample Flow, Sensor Efficiency Battery Condition
- Self-integrity Check: Continuous on-line full self-integrity check eliminates wrong measurements due to Instrument Faults
- Automatic Shut-off: Automatic shut-off of sampling in case of excessive gas Concentration – preserves & prolongs useful life of sensor& Sampling Pump
- USER Friendly: Visual Messages on status on functioning Pleasant Acoustic Buzzer at every change of function & gas displayed
- Display: High quality Alpha Numeric LCD
- Calibration: Automatic Start with fresh air sample
- SPAN Calibration: For 02 automatic at instrument start up. For other gases automatic with calibration gas
- Response Time: Maximum 20 seconds at 95% variation
- Power Supply: (Internal Ni-cad battery pack with external charger) Battery & 22oV/50 Hz
- Working Temperature: -5 to 55°C
- Storage Temperature: 2 -20 to 45°C
- Case: Aluminum
- External Dimension: 170 X 100 X 130 mm
- Weight: 1.5 Kg./cm2
- Probe (Optional): (i) 8mm, Dia SS 600 mm length with 2 meter Hose other length probe for high lens appeal
- Sniffer Probe : To detect & locate the precise position of Gas leak in a pipe network
- Ambient Probe
- Accessories: Leather case with shoulder strap, battery charger, operational manual
Additional Information:
- Item Code: PA-960
Reviews